Quantum Composer Control - Invalid Command Keyword #160

Closed BradSchoenrock closed 5 years ago

BradSchoenrock commented 6 years ago

When trying to set the state of the quantum composer one of several bugs have been short circuited.

This one is the Invalid Command Keyword.


Error -1074000000 occurred at Quantum Composers 9000 Series.lvlib:Utility Write to Instrument.vi

Possible reason(s):

Instrument reports

Error 3: Invalid Command Keyword

Command/Query Sent:*IDN?

Complete call chain: Quantum Composers 9000 Series.lvlib:Utility Write to Instrument.vi Quantum Composers 9000 Series.lvlib:Initialize.vi Quantum Composers 9000 Series Recall-Save Configuration.vi


Good luck future me?

21

stillwer commented 5 years ago

Brad noted that the command that was the problem was: Command/Query Sent:*IDN?. This is a clue that the hardware is not set up properly. I will document the procedure to do so for the future.

The main issue here is that the quantum composer software asks the hardware to identify itself to figure out what features are available. This is done first and needs to be done. Without a response, the software falls on its face. No response will be forthcoming if the port is not set up correctly.

The system has to be set up in combination with the router and the software that comes with the hardware before the labview will run. The failure of the *IDN? command is indicative of that.