Asses Daily Calibration Captures for Herbarium Week 43 2024 #148

Closed chelseagraham closed 1 month ago

chelseagraham commented 1 month ago

WEEK 43

22 October JMJ - WORKHERB0001

Screenshot 2024-10-28 at 11 18 23

Greatest Difference (between center and lower left-hand corner) 11,75 units (11,75/255 = 4,61% error)

This calibration at WORKHERB0001 was within the desired tolerance of 5% and the accepted tolerance of 8% of error.