Open MadsR opened 2 years ago
1.2.12
Ice Nucleation Controller, Temperature Controller
When starting new measurement, temperature controller sometimes peaks to 100%. Starting system at 10°C, gradient close to 0K/min.
Slow and steady increase with minimal overshoot.
Overshot significant.
~5K/min overshoot
Update Gradient calculations appear to be significantly off once system temperature readings start. Investigations should be made to ensure gradient calculations are correctly handled.
Ice-Nucleation-Controller version
1.2.12
Area(s) with issue?
Ice Nucleation Controller, Temperature Controller
Steps to reproduce
When starting new measurement, temperature controller sometimes peaks to 100%. Starting system at 10°C, gradient close to 0K/min.
✔️ Expected Behavior
Slow and steady increase with minimal overshoot.
❌ Actual Behavior
Overshot significant.
~5K/min overshoot