Cooling -> Heating -> Cooling cycle #31

Closed MadsR closed 1 year ago

MadsR commented 2 years ago

Description of the new feature / enhancement

Implement heating functionalities into sampling.

Example temperature:

start: 10°C Cool -> -30°C Heat -> 70°C Cool -> -30°C Heat -> 120°C Cool -> -30°C Heat -> 200°C Cool -> -30°C

Scenario when this would be used?

Periodic healing->cooling->heating experiments.

Supporting information

Structural frame of system is currently 3D printed, which won't survive the high temperatures.

Additionally cold water circulator will overpower peltier elements if circulated resulting in a maximum temperature below 200°C.

Tasks: