Aarhus-University-MPE / Ice-Nucleation-Controller

Software for controlling Ice Nucleation Setup, used to detect ice nucleation activity.
0 stars 0 forks source link

Automatic Non Uniformity Correction Uncertainty Interference #49

Closed MadsR closed 11 months ago

MadsR commented 1 year ago

Ice-Nucleation-Controller version

1.3.0

Area(s) with issue?

Hardware Issue, Other

Steps to reproduce

When doing Uncertainty measurements, automatic Non Uniformity Correction (NUC) interferes with measurements, resulting in large jumps in temperature correction, mid-experiment.

image

✔️ Expected Behavior

No response

❌ Actual Behavior

No response

MadsR commented 1 year ago

Possible Solution 1 Adding a function to disable automatic NUC.

Possible Solution 2 Manually run NUC before experiment start. Still makes it difficult to compare between experiments.

MadsR commented 1 year ago

Implemented in Release v1.3.1 https://github.com/Aarhus-University-MPE/Ice-Nucleation-Controller/releases/tag/v1.3.1

Strategy still to be determined.

MadsR commented 1 year ago

Supporting Material https://www.flir.com/discover/professional-tools/what-is-a-non-uniformity-correction-nuc/

https://www.flir.com/support-center/instruments2/how-do-i-manually-control-the-nuc-auto-calibration-in-the-flir-a35-and-a65/

MadsR commented 11 months ago

NUC impact is added to the uncertainty calculations of the system.

image