o There are two indicators having to do with switch heater. It’s not very
intuitive what they mean.
o Add a warning indicator based on temperature.
o One time: setting the temperature to 5K and turning the needle
valve/heater to “Auto” failed to cause any reaction.
o Re: Stopping reading the cryogen levels.
Reproducible: if we take over local control on the physical front panel,
the Labview program stops reading properly.
Not reproducible: many other random times the Labview program stops
reading properly.
o Needle valve percentage estimate: when going from 35% to 100%, the
estimated % goes to zero, then starts counting up from there.
o When “Go to zero” is used, perhaps we should have the program turn off
the switch heater for us.
o When going to field, it stops reading/recording the temperature
Original issue reported on code.google.com by sbr...@byu.net on 4 Mar 2010 at 5:28