Accuview Spectrophotometer Test (Oct.15-Oct.29) #33

Closed VanessaQi closed 5 years ago

mlee1721 commented 6 years ago

Found issue with Accuview Spectrophotometer: wrong vial was being used for machine. Recalibrated absorbance using DI water. Absorbance larger than 0, which was good. Tested absorbances of various humic acid samples: 5 mg/L, 10 mg/L, 15 mg/L. Data again was plotted, shown below.

image

mlee1721 commented 6 years ago

Measured turbidity of humic acid stream was very low ; suggests that humic acid turbidity may be negligible in overall stream (when mixed with other inputs like clay).

VanessaQi commented 6 years ago

Beer's law is obeyed ---consistent w/ experimental results obtained from the design lab