The reboot method gives us an error when executed, as it receives multiple lines but only expected one. If we add the multiline=True argument it runs without crashing. We are not sure if this is laser specific behavior as we only have one iBeam smart 515nm laser and can't test it on others.
The error:
---------------------------------------------------------------------------
TopticaError Traceback (most recent call last)
Cell In[2], line 1
----> 1 laser.reboot()
File c:\Users\fulapuser\NOMAD-CAMELS\.desertenv\Lib\site-packages\pylablib\devices\Toptica\ibeam.py:100, in TopticaIBeam.reboot(self)
98 def reboot(self):
99 """Reboot the laser system"""
--> 100 self.query("reset system",reply=False,)
File c:\Users\fulapuser\NOMAD-CAMELS\.desertenv\Lib\site-packages\pylablib\devices\Toptica\ibeam.py:92, in TopticaIBeam.query(self, comm, multiline, keep_whitespace, check_error, reply)
90 for i in range(ntries):
91 try:
---> 92 return self._do_query(comm,multiline=multiline,keep_whitespace=keep_whitespace,check_error=check_error,reply=reply)
93 except TopticaError:
94 if i==ntries-1:
File c:\Users\fulapuser\NOMAD-CAMELS\.desertenv\Lib\site-packages\pylablib\devices\Toptica\ibeam.py:85, in TopticaIBeam._do_query(self, comm, multiline, keep_whitespace, check_error, reply)
83 return lines
84 elif len(lines)>1:
---> 85 raise TopticaError("expected single line, but got response {} with {} lines".format("\n".join(lines),len(lines)))
86 return lines[-1] if lines else None
TopticaError: expected single line, but got response SAVE:FSP
iBEAMsmart pulse option - Ultra Compact Diode Laser
(C)2009, TOPTICA Photonics AG
SerialN: iBEAM-SMART-515-S-A3-17694
PowerUP: 00357h
LaserON: 00062h
LDDinit: PASS
TECloop: LOOP
TECpass: WAITUSER
LaserON: USER
Channel: USER
RFoscON: USER
PWRloop: USER
PDreset: USER
ExtModu: USER
SysBoot: DONE with 16 lines
The reboot method gives us an error when executed, as it receives multiple lines but only expected one. If we add the
multiline=True
argument it runs without crashing. We are not sure if this is laser specific behavior as we only have one iBeam smart 515nm laser and can't test it on others.The error: