Closed slaarhuis closed 8 months ago
Happens on 3.0.1 but also on earlier versions
It’s as expected, as there is a buffer. When completely empty the pod will start screaming.
Thx Jon! Will it also scream if the pump sounds are silenced? I did hear the < 10 U sound, but I was able to keep dosing more than 10 units after that.
Thanks (for all the great work!), Stephan
Op di 13 feb 2024 om 22:43 schreef Jon B Mårtensson < @.***>
It’s as expected, as there is a buffer. When completely empty the pod will start screaming.
— Reply to this email directly, view it on GitHub https://github.com/Artificial-Pancreas/iAPS/issues/529#issuecomment-1942668726, or unsubscribe https://github.com/notifications/unsubscribe-auth/AY6TNVRASVTY4FTVI5LYYP3YTPNAPAVCNFSM6AAAAABDHCKT72VHI2DSMVQWIX3LMV43OSLTON2WKQ3PNVWWK3TUHMYTSNBSGY3DQNZSGY . You are receiving this because you authored the thread.Message ID: @.***>
Yes. You can’t turn off the screaming.
I can bolus as much as I want when the indicator of the units left is on 0. I do get a warning when I'm below 10, but after that it keeps bolusing. I guess that should not be the case.