Thanks for the question. Let me add in a few explanations below:
Calculate_Quantitation_Methods
based on a given set of calibration standards provided in the file StandardsConcentrations.csv
. Alternatively, the user can provide these values if they are already known (e.g., at the extreme end if a single point calibration based on a known concentration of internal standard is used).Hopefully, some useful references:
QuantitationMethod
class documentation https://abibuilder.informatik.uni-tuebingen.de/archive/openms/Documentation/nightly/html/classOpenMS_1_1AbsoluteQuantitation.html#detailsPlease let me know if that is helpful (or not).
Thank you for the quick response. I will look into this.
With the release of the new SmartPeak user session feature, I must provide a "quantitation methods" file before I can execute the LCMS_MRM_Unknowns workflow preset. However, I do not quite understand what information should be captured in this file.
From the example datasets, I can infer that the following columns are required:
I would appreciate if you could explain how I should fill out the quantitation methods file for my data. Thank you!