BUG - LOW - AMICS PROCESS - Percentags in "Current Result Standards" list includes deleted segments 3.3.0.1701 #386

Open jkabel opened 1 year ago

jkabel commented 1 year ago

The modal percentages that appear beside minerals in the current result standard appear to include results from deleted segments but totals appear correct in Calculation Tables tab:

image

I used silver paint on some samples I thought were for LA-ICP-MS, deleted the paint, but the silver still shows up as ~3% in the current result standard list.

I'd call this low priority.

knapp-amics commented 1 year ago

@ChiLyMan can you run down and see if this is still happening after the UI update?

ChiLyMan commented 1 year ago

@knapp-amics Just checked on the newest build and this is still the case. The modal percentages in the group list do not reflect deleted particle/fields

knapp-amics commented 1 year ago

Thank you @ChiLyMan we will add it to this list for this bashing but also just note that in the release known issues. @jkabel thanks for the report, @MarkGuo7

GGloy commented 1 year ago

@ChiLyMan I checked in vs1846 - if I delete something the calculation tab will reflect the change immediately, but the group list will only update after a re-classification