Verification of new workflow version: RML v8 #374

Closed idalindegaard closed 1 year ago

idalindegaard commented 1 year ago

Lims Issue

For definitions of headings below see Basic Concepts.

Work Flow: RML v8 Protocol: CG002 Reception Control, CG002 RML Pool samples, CG002 RML Initial QC v3 Step:

Description:

With the new workflow setup where we have two separate sequencing workflows (NovaSeq 6000 and NovaSeq X), we need to create new versions of all workflows that end with sequencing. We will remove protocols NovaSeq v3 and Delivery v1, and include a new automation to queue the samples to an intermediate workflow, Queue samples for sequencing. Following documents New workflow in LIMS and Validation and verification, all new versions of workflows need to be verified.

The verification includes

Since many protocols are straight out copied from the previous version they won't go through extensive testing, we will just check that they perform satisfactorily with regular samples.

Actions:

The new workflow is essentially the same as v6, and consists of three protocols:

CG002 Reception Control: same protocol as in RML v6 CG002 RML Pool samples: same protocol as in RML v6 CG002 RML Initial QC v3: new protocol, same as in v1, except:

Protocol Created in Stage Tested in Stage Created in Production
CG002 Reception Control N/A* Done! 2023-05-23 ARW N/A
CG002 RML Pool samples N/A* Done! 2023-05-23 ARW N/A
CG002 RML Initial QC v3 Done! 2023-05-17 ARW Done! 2023-05-23 ARW 2023-06-12 ARW

*This protocol is the same as in the previous version of the workflow

Testing setup:

Eight samples will be run through the entire workflow. For protocols CG002 Reception Control and CG002 RML Pool samples, the samples will just be passed on through the steps to test that they perform the same as in the current workflow (v6). For protocol CG002 RML Initial QC v3, only the last step has been altered as compared to v1. This step will be verified to see that the new function of queueing samples to the workflow Queue Samples for Sequencing works.

The plate was named verrml230523

All samples were from #327440

LIMS ID (Sample) Sample Name Apptag
ACC11446A1 PB-P-OLVA4006-CFDNA-Ctrl1-KH20230213-C420230214 RMLP15R100
ACC11446A2 PB-P-OLVA4006-CFDNA-Ctrl2-KH20230213-C420230214 RMLP15R100
ACC11446A3 PB-P-OLVA4006-CFDNA-Ctrl3-KH20230213-C420230214 RMLP15R100
ACC11446A4 PB-P-OLVA4006-CFDNA-Wash1-KH20230213-C420230214 RMLP15R100
ACC11446A5 PB-P-OLVA4006-CFDNA-Wash2-KH20230213-C420230214 RMLP15R100
ACC11446A7 PB-P-OLVA4006-CFDNA-evaporation1-KH20230213-C420230214 RMLP15R100
ACC11446A8 PB-P-OLVA4006-CFDNA-Evaporation2-KH20230213-C420230214 RMLP15R100
ACC11446A9 PB-P-OLVA4006-CFDNA-Evaporation3-KH20230213-C420230214 RMLP15R100
Protocol Step Test specification Comments/Results P/F Sign
CG002 Reception Control CG002 - Reception Control NA** All 8 samples passed without error messages. All EPPs functional P 2023-05-23 ARW
CG002 RML Pool samples CG002 - RML Library Pooling NA** All 8 samples passed without error messages. All EPPs functional P 2023-05-23 ARW
CG002 RML Initial QC v3 CG002 - Qubit QC RML NA** All 8 samples passed without error messages. All EPPs functional. Used Qubit values from original run on 230216 (320 bp, 1,78 ng/ul) P 2023-05-23 ARW
CG002 RML Initial QC v3 CG002 - Aggregate QC (DNA) Test using 8 samples in one pool. Check that there are no errors with EPPs. Check that samples are queued to the correct workflow All 8 samples passed without error messages. All EPPs functional. The pool was successfully aggregated and queued to the correct workflow P 2023-05-23 ARW

**The step is the same as in the previous version of the workflow and will not be tested specifically with different scenarios

Project resources:

Screen shots of all steps can be found on Google Drive

Conclusions:

The workflow passed the verification and is ready to be used in production.

Sign off

Head of Unit Lab Date
ALY 2023-08-25

Implementation

What is needed to implement changes Item Date signature
Information to staff 1. Email sent out to email adress clinical_production@scilifelab.se
2. Inform during lab meeting
3. Inform lab staff on Slack when workflow is activated
1. 2023-08-23 IL
2. 2023-08-28 ARW + IL
3. 230829/ARW+IL
Documentation No Atlas updates needed, no mention of protocol version in NovaSeq 6000 Sequencing method N/A
Test in production Monitor workflow performance during the first week after activation 2023-09-05 ARW+IL
Move negative controls N/A N/A
Move ongoing samples N/A (workflow update was performed between two NIPT runs) N/A
Archive old workflow version and hide from lab view Archive workflow RML v7 KSV
amandaraud commented 1 year ago

Testing in stage done! Ready to be created in production.

amandaraud commented 1 year ago

Workflow created in production (pending), ready to be activated and tested.

amandaraud commented 1 year ago

Ready to be signed off @annaengstrom