TestReport: Patient-Connection Port specs #118

Closed CohenLabPrinceton closed 4 years ago

CohenLabPrinceton commented 4 years ago

a) a female 15 mm conical connector complying with ISO 5356‐1:2015, or

b) a coaxial 15 mm/22 mm conical connector complying with ISO 5356‐1:2015.

CohenLabPrinceton commented 4 years ago

I think we use the co-ax 15/22, yes? @juliennelachance