TestReport: Testing Error Reporting #87

Closed ProfDJCohen closed 4 years ago

ProfDJCohen commented 4 years ago

Ref: CR501 201.5.101.3

VENTILATORY SUPPORT EQUIPMENT testing errors

ProfDJCohen commented 4 years ago

@zajdel @CommReteris @sneakers-the-rat @Sophie We need to make sure that we have a table of measurement uncertainty, etc. that we report in our labeling parameters and for every measured/derived parameter.

sneakers-the-rat commented 4 years ago

should the UI display static +/- % error tolerance message next to measured parameters?

CohenLabPrinceton commented 4 years ago

Nope. These need to go in the labeling documentation and disclosures in the manual, not the machine UI @sneakers-the-rat

CohenLabPrinceton commented 4 years ago

Update: @zajdel will test this using testing conditions for Pressure Control table from the ISO table. @SophieDvali can help fit the data if needed.

SophieDvali commented 4 years ago

I'm a little confused about the wording here, does this just mean we have to report the uncertainty of all measured quantities? Is this separate from the EUA items for reporting the error of the individual measured quantities?

CohenLabPrinceton commented 4 years ago

@SophieDvali these are the same things. We need to report the error for any quantities we show the user. I think if you have the error values for flow and volume then we should be good.

sneakers-the-rat commented 4 years ago

unassigning myself since this is a docs thing for the hardware team. lmk if you need help using the docs tho.