EDAX/TSL x pattern center reversed in simulated patterns #161

Closed Habakuk224 closed 2 years ago

Habakuk224 commented 2 years ago

Hello,

I am trying to reindex a material using DI. However, the DI result is approximately 5° off the original data indexed by Hough indexing in OIM Analysis. After testing of various processing steps, I have discovered, that the pattern center defined in DItutorial for EDAX systems is probably reversed in x direction. An example follows:

A 120x120 simulation was done by EMEBSD for Euler angles 90, 22, 0 using

xpc = 30
ypc = 0
L = 26100

for BCC material, i.e. the [100] axis should be perpendicular to the detector. The xpc=30 is a bit extreme, but the error is then easily seen (while for typical xpc close to 0, the effect is barely noticeable). x30_y0_E90-22-0 According to the definition from the DItutorial, this corresponds to

x* = 75 %
y* = 50 %
z* = 81.86 %

However, when viewed in OIM Analysis (v8) the pattern center defined in such way is shifted from the actual pattern center position on the [100] axis: nok

Correct pattern center is shown for x* = 25 % ok

Y pattern center is behaving correctly and ypc corresponds to its y* counterpart.

marcdegraef commented 2 years ago

Hi,

this was documented for EMsoft 5.0 on https://github.com/EMsoft-org/EMsoft about half-way down the page:

*

      *For Dictionary Indexing (DI) users, we have changed our
      convention for EBSD pattern orientation to be the same as the
      vendor convention; i.e., EBSD patterns are displayed looking
      from the detector to the sample. The practical consequence is
      that for older DI runs the sign of the xpc (x-component of
      pattern center) will need to be reversed when used with the
      Release 5.0 indexing programs.*

Regards, Marc.

On 6/3/22 12:11 PM, Habakuk224 wrote:

Hello,

I am trying to reindex a material using DI. However, the DI result is approximately 5° off the original data indexed by Hough indexing in OIM Analysis. After testing of various processing steps, I have discovered, that the pattern center defined in DItutorial https://github.com/EMsoft-org/EMsoft/wiki/DItutorial#51-setting-the-detector-geometry for EDAX systems is probably reversed in x direction. An example follows:

A 120x120 simulation was done by /EMEBSD/ for Euler angles 90, 22, 0 using

|xpc = 30 ypc = 0 L = 26100 |

for BCC material, i.e. the [100] axis should be perpendicular to the detector. The xpc=30 is a bit extreme, but the error is then easily seen (while for typical xpc close to 0, the effect is barely noticeable). x30_y0_E90-22-0 https://user-images.githubusercontent.com/72706923/171902105-758ec3b8-2d32-4818-af40-aaea7d5ec8e3.png According to the definition from the DItutorial, this corresponds to

|x = 75 % y = 50 % z* = 81.86 % |

However, when viewed in OIM Analysis (v8) the pattern center defined in such way is shifted from the actual pattern center position on the [100] axis: nok https://user-images.githubusercontent.com/72706923/171902134-7267de64-0063-4c85-891d-929ac56964cf.png

Correct pattern center is shown for x* = 25 % ok https://user-images.githubusercontent.com/72706923/171902128-1dc4b9b1-8612-4181-9b0b-2fd36b820dfe.png

Y pattern center is behaving correctly and ypc corresponds to its y* counterpart.

— Reply to this email directly, view it on GitHub https://github.com/EMsoft-org/EMsoft/issues/161, or unsubscribe https://github.com/notifications/unsubscribe-auth/AB26VWEXDFXYIOA5XSSNQH3VNIVDBANCNFSM5XZPIIXA. You are receiving this because you are subscribed to this thread.Message ID: @.***>

Habakuk224 commented 2 years ago

My mistake, thank you for clarification.

Regards, Jiri Kozlik