GSTT-CSC / hazen

Quality assurance framework for Magnetic Resonance Imaging
https://github.com/GSTT-CSC/hazen
Apache License 2.0
22 stars 12 forks source link

Slice position only measuring on slice 11 #446

Open RebeccaThornley opened 2 months ago

RebeccaThornley commented 2 months ago

Describe the bug From a recent QA on a Siemens Sola, we found that the slice position test only gave a result for slice 11 and not slice 1. Screenshot below and data set attached.

Note that all other tests worked ok, and slice position test worked for other orientations.

image

Desktop (please complete the following information):

ACR_COR_11001.zip

RebeccaThornley commented 3 weeks ago

This happened again on a QA of a Siemens Aera. I am attaching the data sets.

Body_tra_PSN.zip Body_sag_PSN.zip