skin dose in SAXS #12

Closed jmholton closed 11 months ago

jmholton commented 3 years ago

Just did a run of water in a cylindrical quartz capillary. Dose to the "skin" of the sample was low, presumably due to photoelectron escape. However, there should be more photoelectrons coming into the water from the quartz than the other way round, so in reality the skin dose should be higher. This is a bug, right?

jdickerson95 commented 3 years ago

Hi James, I've added photoelectron entry from the surrounding into SAXS so it should now be fixed.