Closed martin-machacek closed 8 months ago

martin-machacek commented 8 months ago

Everytime I turn the device on, it is in the Hold mode of Instantaneous screen. It does not matter if I previously turned it off in Hold mode or not. While turning it on, I release the center OK key as soon as the LED turns off and Rad Pro splashscreen appears.

martin-machacek commented 8 months ago

I may be holding the key for too long. If I release the key sooner, it still turns on, but not in Hold mode.

Gissio commented 8 months ago

I'm glad you could install Rad Pro successfully on your device!

Regarding the issue, this can be fixed! Thanks for the report.

Tig3rch3n commented 8 months ago

Just confirmed the issue. If you hold Power just 1 Sec, Instantaneous is NOT on Hold If you hold Power for >2 Sec it starts on Hold

Gissio commented 8 months ago

Beta20 should fix this. Let me know.

martin-machacek commented 8 months ago

I can confirm that it is fixed 👍

Gissio commented 8 months ago

Nice!