Closed Hendricks-Laboratory closed 1 month ago
If we change all of these 55s to 50s I think we'd probably be okay!
The tip is still going too far in. I think we need to change the 55 on line 459 to 50 and try that, but then we might need to adjust both to something even smaller (although the tips are 59-60mm long, so I don't know why the 55 isn't working)
Did this change!
I'm going to leave this open until it's tested
This is just barely good enough, with surface tension it's not going above the tip, but with a different solution/temperature/luck, it probably would touch. Let's change both of these 50s to 48 and I think we'll be good!
During mixing a 15/50 mL tube that's mostly full, the maximum depth the tip should go down is based on the length of the tip (so that we only submerge the tip and not the pipette itself). It seems like this is working, but it's going just a little too deep so it's just submerging past the tip/touching the pipette. I think 5 mm less deep/higher would solve the problem and then I can test again. Overall mixes are looking really good though!