RegisteredContourerScenario2 - I #10

Closed GoogleCodeExporter closed 9 years ago

GoogleCodeExporter commented 9 years ago
What version of the product are you using? On what operating system?
MMRO-II v1.2

Please provide any additional information below.

Used the Hybrid Image Set with CT and MR on the same FOR.

Class UIDs:
1.2.840.10008.5.1.4.1.1.2: CT Image Storage
1.2.840.10008.5.1.4.1.1.4: MR Image Storage

Test Run 1
2013RegisteredContourerScenario2, CT-PET, Structure on PET

Validate: Selected Dataset definition: "RT Structure Set Storage SOP Class". 
Using AE Name "DICOM" - AE Version "3.0" 
IHE-RO Error : Series Instance UID Stucture Set Rule - 'SeriesInstanceUID' 
(0x0020000E) of the Structure Set does not match the 'SeriesInstanceUID' of the 
CT-images
Link to Detailed Result 

IHE-RO Error : Contour Image Sequence Structure Set Rule - 
'ContourImageSequenceTag' (0x30060016) does not have a reference to every 
CT-Images, reference to 100 CT-images, 221 CT-Images should be referenced
Link to Detailed Result 

IHE-RO Error : Contour Image Sequence Contour Interrelation Rule - Number of 
Image Sequence Items is one but zcoordinates of the contour does not match the 
z-coordinate of Image Position Patient in the Image.
Link to Detailed Result 
IHE-RO Error : Referenced SOP Instance UID Contour Module Rule - 
'ReferencedSOPInstanceUID' (0x00081155) has no reference to 'SOPInstanceUID' 
(0x00080018) of a CT-Images
Link to Detailed Result 

IHE-RO Error : Contour Data Rule - Number of Image Sequence Items is one but 
zcoordinates of the contour does NOT match the z-coordinate of Image Position 
Patient in the Image.
Link to Detailed Result 

Test Run 2
As first attempt, a Structure Set was created on the CT. The test tools 
reported the error below. This may indicate, that an MR is expected.

Validate: Selected Dataset definition: "RT Structure Set Storage SOP Class". 
Using AE Name "DICOM" - AE Version "3.0" 
IHE-RO Error : Value Rule - The attribute 'ContourReferencedSOPClassUID' 
(0x30060039[1]/0x30060040[1]/0x30060016[1]/0x00081150) contains the value: 
1.2.840.10008.5.1.4.1.1.2 but should contain the value: 
1.2.840.10008.5.1.4.1.1.4. 

As second attempt, a Structure Set was created on the MR. The test tools 
reported the error below. The tool complains about CT references, which should 
not be requested (since CT anyways are not expected – see above).
Further on, the Series Instance UID of a CT can never be the same as the Series 
Instance UID of a Structure Set, since both must belong to different series.

Validate: Selected Dataset definition: "RT Structure Set Storage SOP Class". 
Using AE Name "DICOM" - AE Version "3.0" 
IHE-RO Error : Series Instance UID Stucture Set Rule - 'SeriesInstanceUID' 
(0x0020000E) of the Structure Set does not match the 'SeriesInstanceUID' of the 
CT-images
Link to Detailed Result 

IHE-RO Error : Contour Image Sequence Structure Set Rule - 
'ContourImageSequenceTag' (0x30060016) does not have a reference to every 
CT-Images, reference to 200 CT-images, 221 CT-Images should be referenced
Link to Detailed Result 

IHE-RO Error : Contour Image Sequence Contour Interrelation Rule - Number of 
Image Sequence Items is one but zcoordinates of the contour does not match the 
z-coordinate of Image Position Patient in the Image.
Link to Detailed Result 

IHE-RO Error : Referenced SOP Instance UID Contour Module Rule - 
'ReferencedSOPInstanceUID' (0x00081155) has no reference to 'SOPInstanceUID' 
(0x00080018) of a CT-Images
Link to Detailed Result 

IHE-RO Error : Contour Data Rule - Number of Image Sequence Items is one but 
zcoordinates of the contour does NOT match the z-coordinate of Image Position 
Patient in the Image.
Link to Detailed Result 

Original issue reported on code.google.com by woutervr...@gmail.com on 5 Dec 2014 at 7:27

Attachments:

GoogleCodeExporter commented 9 years ago
Series Instance UID Stucture Set Rule is updated.
Contour Image Sequence Structure Set Rule is updated.
Contour Image Sequence Contour Interrelation Rule is updated.
Referenced SOP Instance UID Contour Module Rule is updated.
Contour Data Rule is updated.

Original comment by woutervr...@gmail.com on 9 Jan 2015 at 10:30

GoogleCodeExporter commented 9 years ago
Fixed in version 1.3, downloadable on 
http://ihe-ro.org/doku.php?id=doc:testtools:test_tools 

Original comment by woutervr...@gmail.com on 27 Jan 2015 at 2:16