RT structure set validation fails unless the only slice in the contour is the last slice of the image #60

Open KatherinePapcun opened 8 years ago

KatherinePapcun commented 8 years ago

I have encountered this on several of the scenarios, including the MMRO-II registered contouring scenario and the BRTO contourer scenario 1.

There are two failures listed:

"IHE-RO Error: Contour Image Sequence Contour Interrelation Rule - the 0x00080018 (Sop Instance UID) with value [the last SOP of the image] does not match the 0x00081155 (Referenced SOP Instance UID) value: [the first SOP that is included in the contour data]"

"IHE-RO Error: Contour Data Rule - the 0x00080018 (Sop Instance UID) with value [the last SOP of the image] does not match the 0x00081155 (Referenced SOP Instance UID) value: [another SOP that is included in the contour data]"

Creating a contour that only occupies the last slice of the image seems to satisfy the test, but that hardly seems like the desired behavior.

iherotesttools commented 8 years ago

Updated rules