Closed kjwoodsISIS closed 1 month ago
From the VI it looks like they are just using the SR850 to change the frequency, so I may put a training ticket on this
Required for WISH student project June/July 2024, and then for experiment Autumn 2024.
MVP requires logging of the frequency (as mentioned above), but setting of basic parameters (unknown at this stage) desirable.
UPDATE - see below for commands. Suggest replicating VI functionality as is relatively simple, rather than MVP.
VI front panel:
Brief analysis of VI:
<CR>
or \r
or 0x0D
FREQ
command (the only write command used)FREQ?
OUTP ? x
, where x is 1,2,3,4 respectivelyHave located unit. Team WISH double-checking it's the correct one.
The IOC and corresponding OPI has been created:
IOC Support: https://github.com/ISISComputingGroup/EPICS-SR850/pull/1 EPIC IOC: https://github.com/ISISComputingGroup/EPICS-ioc/pull/861 GUI/OPI: https://github.com/ISISComputingGroup/ibex_gui/pull/1710 Release Notes: https://github.com/ISISComputingGroup/IBEX/pull/8385 EPICS Top: https://github.com/ISISComputingGroup/EPICS/pull/354 wiki: SR850 manual: \isis\shares\ISIS_Experiment_Controls\Manuals\Stanford_Research_SR850_Lock-In_Amplifier
As a WISH scientist I want to be able to control a Stanford RS SR850 Lock-In Amplifier in my experiments.
Acceptance Criteria
Notes
C:\LabVIEW Modules\Drivers\Stanford SR850
).Time in Planning
01:08:47 02/05/24