Z-stage seems to initiate in wrong position #50

Closed wl-stepp closed 8 months ago

wl-stepp commented 8 months ago

Could be that there is a voltage remaining on the DAQ due to the aborted acquisition

wl-stepp commented 8 months ago

It was actually the stage position that was very far on the edges. Thus it could not move all the way. Resetting it to the center fixed that. I removed the resetting of the position on close. This sometimes doesn't get called if we crash. This should be put in something like a finally block. Not sure if it would be enough in a python finally block, as sometimes we have to crash out hard. Maybe as an option on shutdown or something like that?