MassBank / MassBank2NIST

A converter between MassBank and NIST (and Bruker .library) formats and back
2 stars 2 forks source link

Tags for single CE, ramped CE and bbCID #16

Open schymane opened 10 years ago

schymane commented 10 years ago

Current specs can deal with single CE or ramped CE: AC$MASS_SPECTROMETRY: COLLISION_ENERGY 20 kV AC$MASS_SPECTROMETRY: COLLISION_ENERGY Ramp 10-50 kV

Options for broadband (MSe, DIA ...)

schymane commented 10 years ago

MS$FOCUSED_ION: ISOLATION_WINDOW MS$MASS_SPECTROMETRY: MS_TYPE: MS_ALL

schymane commented 10 years ago

Optional: add comment field to say bbCID etc...

schymane commented 10 years ago

Answer from Bruker: The entry MassRange: 44.9801,1005.2 is equivalent to the isolation window for bbCID No entry for isolation width per se, "Usually we tune the instrument to an isolation width to capture all isotopes of a precursor. That’s why we report only one value – the mono-isotopic mass". To be included in a future version.