Humidity sensor's dew point temperature seems wrong #262

Closed stillwer closed 5 years ago

stillwer commented 5 years ago

The dew point temperature is reporting higher than the actual temperature when the humidity raises. That can not be right.

stillwer commented 5 years ago

Posted this problem to the Servers Check forum and got a response that someone else had the same problem that they solved by reverting the firmware from V8 to V7.5. Will try that now.

stillwer commented 5 years ago

The temperature response is better but does not match a simple calculation I found online.

stillwer commented 5 years ago

Just tested moving the sensor from the wood platform that Scott had to the metal floor of the container (Oct. 23, 2019 at ~15:30 UTC on MPD 2). The temperature dropped (also because I moved it away from the computer fans) but the dew point is not going crazy. Seems like mounting to metal is okay...just need to do so.

stillwer commented 5 years ago

Sensor is mounted to MPD2's metal frame and seems to be giving reasonable results. The dew point is still a bit weird but the relative humidity seems reasonable. I will mark this as resolved for now more or less because we don't use that number and it is irrelevant to our operations.