ICP Torch went out unexpectedly. #2

Closed faytrow closed 3 years ago

faytrow commented 3 years ago

During normal ICP operation, the plasma/torch went out and gave an E9006 error indicating the 'RF Reported over current failure'.

Upon attempting to re-ignite the plasma, the torch would light for about 5-10 seconds, then flicker slightly before puffing out. ITEVA reported an E9000 error and an E1133 error, neither of which are informative beyond the plasma going out.

faytrow commented 3 years ago

Update:

We replaced the torch and the sample intake tube, and replaced the Argon tank. This extended the duration the torch stayed lit, but not by much nor solved the errors.

faytrow commented 3 years ago

Solved:

We filled the rinse cup back up with fluid, and the plasma now stays on.

This did give a 'Debug Wavelength Check' warning, but that was solved by re-aligning the torch with a 2 ppm zinc standard