Closed faytrow closed 3 years ago
Update:
We replaced the torch and the sample intake tube, and replaced the Argon tank. This extended the duration the torch stayed lit, but not by much nor solved the errors.
Solved:
We filled the rinse cup back up with fluid, and the plasma now stays on.
This did give a 'Debug Wavelength Check' warning, but that was solved by re-aligning the torch with a 2 ppm zinc standard
During normal ICP operation, the plasma/torch went out and gave an E9006 error indicating the 'RF Reported over current failure'.
Upon attempting to re-ignite the plasma, the torch would light for about 5-10 seconds, then flicker slightly before puffing out. ITEVA reported an E9000 error and an E1133 error, neither of which are informative beyond the plasma going out.