Replaced Outer Cocktail Valve #22

Open jirhiker opened 7 years ago

jirhiker commented 7 years ago

Starting week of June 6 cocktail blank progessively increased. Matt H. noticed increase when blank was ~18fA.

A high dynamic rise rate in the mass spec was observed when outer cocktail valve (W) was CLOSED lead us to infer a small leak in the valve bellows.

W has replaced 6/12/17 with a new valve. Minibone backed for 4hours at 175C

Following bake, actuating W still releasing small but significant amount of gas.

Cycling W appears to help clean it up. W cycled 1400 times using a Procedures pyscript (Actuate_w.py on Felix)