Replaced inner pipette (Z) tip #26

Open jirhiker opened 5 years ago

jirhiker commented 5 years ago

inner air pipette valve was conducting. at first diagnosed as leaking bellows on Y. but further investigation with helium revealed that Z was actually conducting. Most evidence at first pointed to leaking bellows on Y (blank was greater than a normal air shot) but a few tests conclusively proved that Y was leak tight and Z was conducting.

replaced valve driver for manual valve above air tank