OpenHantek6022 is a DSO software for Hantek USB digital signal oscilloscopes 6022BE / BL. Development OS is Debian Linux, but the program also works on FreeBSD, MacOS, RaspberryPi and Windows. No support for non-Linux related issues unless a volunteer steps in!
Hi Ho-Ro,
please see attached screenshot below - Probe connected to calibration out, frequency set to 20kHz.
"Measure" button activated. Markers set at the borders of a single signal period, i.e. 50µs.
Everithing (except the frequency in the right upper corner) is calculated / displayed correctly,
see green boxes in the screenshot. The red boxed marker frequency is IMHO wrong - 5kHz.
Shouldn't it be 20 kHz ?
If I move both markers to get measured time period for instance of xy.z µs , the marker
frequency follows the markers allways showing x.yz kHz .... 45.6µs yields 4.56 kHz ...
basically time [µs] / 10.00 --> freq [kHz]
Computer environment:
OS: Xubuntu 19.09 amd64 running on VirtualBox 6.1.4 with Guest Additions 6.1.4
Hi Ho-Ro, please see attached screenshot below - Probe connected to calibration out, frequency set to 20kHz. "Measure" button activated. Markers set at the borders of a single signal period, i.e. 50µs. Everithing (except the frequency in the right upper corner) is calculated / displayed correctly, see green boxes in the screenshot. The red boxed marker frequency is IMHO wrong - 5kHz. Shouldn't it be 20 kHz ?
If I move both markers to get measured time period for instance of xy.z µs , the marker frequency follows the markers allways showing x.yz kHz .... 45.6µs yields 4.56 kHz ... basically time [µs] / 10.00 --> freq [kHz]
Computer environment:
Scope device:
MfG Pedro