Closed emmastrand closed 1 year ago
There are an additional 2 Traceable Temperature Probes that have the following issues:
Item serial #: not readable Cat. #: not readable
June 2022: two temperature probes have corrosion on the USB connector and fail to connect to a working thermometer. LZ: I used a kimwipe with EtOH on it to wipe the connector and plugged it into a working thermometer.
Whoops didn't realize this was already closed! See https://github.com/Putnam-Lab/Lab_Management/issues/40.
Item serial #: 192676777 Cat. #: 6412
June 2022: traceable temperature probe reads HHH indicating an extremely high temperature when functioning probe is plugged in. LZ: I attempted to connect probe known to be functional. I also replaced the battery, cleared the memory and blew dust out of the connecting port.