Closed Wmdhampton closed 9 years ago
A few questions...
The DAQ card that I am using is the NI PCI-6052E: http://sine.ni.com/nips/cds/view/p/lang/en/nid/2601 on the BNC-2090 there are switches that can be set to SE or DIFF. The problem occurs in both settings. When I disconnect the DAQ Analogy Output from the Axoclamp the problem goes away. This occurs on channel 0 and 1 for the output from the BNC-2090. However, it does not occur for any of the input channels.
This is now resolved.
The problem is hardware-based (DAQ appears to be malfunctioning).
When ever I connect my:
DAQ Analog Output 0 --> Axoclamp EXT. ME1 COMMAND
My I(nA) display reads that I am injecting 0.6(nA) even though I am not. This goes away as soon I unplug the DAQ Analog Output 0 on my BNC-2090. I though this might be a strange ground loop error. However, this does not occur when the Axoclamp EXT. ME1 COMMAND is connected to a different computer through a Digidata 1440A. Both of my computers are grounded through the same circuit breaker. Any thoughts would be greatly appreciated.