RobertsLab / 2017-hackweek

0 stars 0 forks source link

Assess Sensophone Alarm system #1

Closed sr320 closed 7 years ago

kubu4 commented 7 years ago

Assessed.

Fixed Alarm 3 (Revco - Horner-Devine Lab freezer) to trigger correctly (wires were wired incorrectly causing alarm status to trigger once freezer returned to normal operating temperature range - instead of triggering when freezer temps were outside of normal operating range).

Updated/corrected/improved Sensaphone documentation in LabDocs Wiki.

kubu4 commented 7 years ago

Revco freezer seems to be solely contain Young Lab stuff. Should we add them to Sensaphone call list or should we remove Revco from our Sensaphone and let them deal with monitoring that freezer?

kubu4 commented 7 years ago

Revco should be considered as a backup freezer - not full time use for Young Lab (or Roberts Lab).