Closed grace-ac closed 7 years ago
Opp, I think you're following an old version of the protocol! Repeat Step 2d (I revised it yesterday so you wouldn't encounter your specific issue).
@grace-ac Did that ^ work?
I haven't tried it yet - I'll be back in lab on Monday afternoon to test it out! sorry if you're waiting on it :(
No worries! Just wanted to make sure I wasn't missing anything.
This happened after re-doing 2d:
That's odd! Didn't happen to me on the same machine you're using when I tried it out. What happens when you press cancel? I don't understand what error it's seeing. Any suggestions @emmats ?
I just pressed cancel, and my screen looks exactly how it does in the original screenshot of this issue
Huh...what does it look like when you press OK?
It seems like you kept all of the previously imported transitions before you redid 2d! Try deleting absolutely everything from that lefthand window and re-importing the transition list.
Let me know if you still need my input.
I restarted from the beginning and still got the error.
Things look much better though, after clicking, "OK" and I'm currently working on Step 3b.
@emmats do you know what could be causing this error? I exported the transition list from Skyline itself.
@grace-ac You should open up the transition list and make sure the transitions in your Skyline document are the same ones in the transition list you're importing.
I went through and I matched the numbers on the list to those in Skyline doc. I just highlighted the first three to show what numbers I was matching.
All of the ones in Skyline Document had a match in the original document. The only ones that weren't in the Skyline doc were the ones that were labeled "peptides" rather than a number.
AH those are the PRTC peptides! I think I know what's causing the error so I can fix that for you. It won't affect any of your downstream analyses.
You can also click on the plus signs to expand the protein and see the peptides/transitions associated with it
Okay - I clicked the plus signs and they all match! woo!
Let me know when you fix whatever is causing the error and i'll re-do it!
In the transition list, what would happen if you removed the space between PRTC and peptides, or got rid of that phrase entirely (keeping the other content in that row)?
I tried both suggestions, and both resulted in the same error.
I've never seen this error before so I can't really weigh in on what is going wrong. If you can't figure it out, I can probably stop by tomorrow afternoon or Monday.
I realize that you've been gone @yaaminiv , but I tried it again, because earlier in this thread you said you were going to fix something - thought I'd see if anything has changed.
Error happened again.
Would it have anything to do with the fact that I'm using Excel?
What is the full error message from the screenshot @grace-ac posted in this comment: https://github.com/RobertsLab/project-oyster-oa/issues/27#issuecomment-337018610
I just updated my comment @kubu4 is that what you are looking for?
Nope. The error messages are truncated and can't be read. Need to be able to read the entire message for each line.
I don't think it's an Excel issue! I think it's some sort of problem with Skyline where it can't recognize the peptides from the transition list because the retention times in the transition list do not align with what's in the library...? I'm still working on it, but this issue won't affect anything in the SRM protocol since I haven't dealt with my PRTC peptides yet.
That's as much as I could get.
Not really sure I understand what's happening - whether I should continue with the protocol or wait for you to investigate things @yaaminiv
@grace-ac Continue!! It's possibly over my head as well so don't wait for me.
For kicks @grace-ac , when that line is highlighted, can you copy it (right-click or ctrl-c), and then paste in a text editor for viewing?
@kubu4 No, but I could hold the mouse over and read the remaining words: "match the tolerance on the Instrument tab of the Transition Settings."
Same for all of them.
@grace-ac Hold up! Continue with 2e, then retry 2d. I have a hunch that after changing your transition settings, you'll be able to import everything
Same error. The transition settings were already set according to your protocol from my previous attempts
Darn. I'm not sure what it is so I'll play around with it tomorrow, but again, it won't affect anything else in that protocol
Cool beans - I shall continue!
3b is where I'm stuck - which is where I believe I started this whole issue in the first place.
Does not look like your screenshot for Step 3b:
Ah! Instead of clicking on the protein name, you want to click on the peptide (i.e. the things with the green dots next to them)
So, am I comparing the peptide retention time to the average SRM retention time?
Yup! You compare the average retention time with the retention times in the Skyline document. The correct peak will have all three transition fragments.
@grace-ac Do you think you can create a new issue with the Skyline error message you keep getting? I want to address the other questions in this issue and close it out, then refer emma to the new issue you'll create!
Resolved issues at 3b, but not the import issue at 2d. @grace-ac will create a new issue and I will continue the discussion there.
So, everything has worked great- I had no issue finding the RAW files in the downloads "Archive" folder today.
However, I'm at Step 3b: Clean Dilution Curve Data, and I don't know what is wrong. All the windows are blank...?