The FP-SNS-DATALOG2 function pack represents an evolution of FP-SNS-DATALOG1 and provides a comprehensive solution for saving data from any combination of sensors and microphones configured up to the maximum sampling rate. Please check st.com where a more recent version of the software might be available.
https://www.st.com/en/embedded-software/fp-sns-datalog2.html
Other
16 stars 1 forks source link

hsdatalog_GUI does not store the device configuration on the SD card (STWIN.box). #17

Closed Ferag-AG closed 2 months ago

Ferag-AG commented 2 months ago

Hi

Uses version 2.1.1. Hardware: STWIN.box

In the HSDPython_SDK hsdatalog_GUI example, when I modify the device configuration and store it on both the computer and the device, the device_config.json file on the computer is updated correctly, but the device_config.json file on the SD card is not.

I hope you'll be able to check this, because if it doesn't work as expected, and the STWIN.box couldn't be configured as intended for measurement recording, this may lead to having to redo a whole set of measurements.

Thanks in advance.

pradolis commented 2 months ago

Hello @Ferag-AG Added in pipe for v2.2.0, planned for end of July.

Best regards Simone

pradolis commented 2 months ago

Hello @Ferag-AG I confirm the bug on the FW side. Fixed in upcoming v2.2.0. I will keep the issue open and notify you once v2.2.0 is online. Then, I'll close the issue.

Best regards Simone

Ferag-AG commented 2 months ago

Hello @pradolis

I really appreciate your reactivity and thank you for your efforts.

Kind regards Ludwig

pradolis commented 2 months ago

Hello @Ferag-AG I've just pushed the v2.2.0.