Open wdwvt1 opened 4 years ago
e.g. adding this from Qi and Amita 1.) Turn of all modules on screen. 2.) Open mass spec cover (temp may be hot). Wipe with EtOh spray bottle above mass spec. 3.) Close MassHunter. Remove Mass Hunter Process (from Windows search. 4.) Turn off IP-LC pump, autosampler, column. 5.) Disconnect "plug 2" from I-P LC and replug into "normal LC" 6.) Disconnect Plug 1 from I-P LC and replug into "normal LC" 7.) Remove wire from IP LC to MS and reconnect the other "normal LC" to the MS. 8.) Turn on "normal LC" pump, autosampler, column. Adjust waste tubes if needed. 9.) Wash for 6-8 hrs with 40% H20, 50% isopropanol, 10% acetic acid using ion pair cleaning method 20190314
A couple to-dos:
6470
procedure here