Open berten76 opened 7 years ago
xDrip / xBridge will need 2 readings from "the source transmitter" before you can calibrate it. Since each transmission happens every 5 mins, you should be able to calibrate xDrip after 10 mins. If you have old blood readings, you will need to redo the poke and get two readings with two testing strips. The testing can happen one after another, after which you input your two numbers into xDrip.
After starting a sensor, XDrip+ ask me for a double calibration. but i cant enter the time i took it. How long should it be between the first blood test and the second blood test. How does xdrip understands how long the timespan was? (im usinig libre). I have read there are calibration plugins, where can i find these?
Thanks Christofer