Closed tdowrick closed 3 years ago
In GitLab by @ThomasDowrick on Nov 6, 2018, 09:13
changed title from Longer default time for Pivot/Spin calibration to Longer default time for Pivot/Spin calibration{+ (10 seconds or more)+}
In GitLab by @ThomasDowrick on Nov 23, 2018, 09:29
changed the description
In GitLab by @JonathanShapey on Dec 14, 2018, 16:56
Completed
In GitLab by @JonathanShapey on Dec 14, 2018, 16:56
closed
In GitLab by @ThomasDowrick on Oct 31, 2018, 14:59
We are using the pivot/spin calibraiton module from SlicerIGT to calibrate the probe.
The default data collection time is 5 seconds, but in our experience 10+ seconds is needed for success, otherwise the error is to high, or not enough data is collected.
The default value should be changed programatically in the Slicelet.