Deviation in values of E are as follows for the LUT implemented in ShapeOut for a 20 um channel:
and for an extended version of the LUT (also in a 20 um channel):
The mistake affects small cells the most. If the object diameter is less than 70 % of the channel size the values of E were underestimated by from about 5 % and up to 60 % for much smaller cells.
The new and correct LUT is this:
For a 20 um channel it covers the following region (values are calculated for a flow rate of 0.04 ul/s and a viscosity of 15 mPa s
The isoelasticity lines retrieved from this LUT are the following:
See also: https://github.com/ZELLMECHANIK-DRESDEN/dclab/issues/39
Due to a mistake in the design of the underlying numerical simulations, the Young's modulus computed from cell area and deformation, as currently implemented in ShapeOut/dclab, is wrong. As a result, the Young's modulus of small cells is underestimated.
Currently, new simulations are underway and will be implemented in ShapeOut/dclab as soon as possible.