Manual Stim doesn't stop and causes NR to crash #13

Open GoogleCodeExporter opened 9 years ago

GoogleCodeExporter commented 9 years ago
What steps will reproduce the problem?
1. Under the Stim/Output tab, in the Manual Stim section, click start (with any 
parameters)
2. Click Stop
3. Doesn't stop, kaboom

What is the expected output? What do you see instead?
The stop button should be pressed, stimulation should stop, and NR should 
respond to commands. Unfortunately, it doesn't. I'm not sure it's actually 
stimulating in the first place (not seeing anything with a voltmeter).

What version of the product are you using? On what operating system?
NR v0.6.0.0, Vista

Please provide any additional information below.
I played around with some of the hardware settings (General AO, pulsatile stim 
time), but none of them did anything.

Original issue reported on code.google.com by nlaxpat....@gmail.com on 12 Jul 2011 at 8:43

GoogleCodeExporter commented 9 years ago
[deleted comment]
GoogleCodeExporter commented 9 years ago
More information is needed on the nature of this crash. Is any exception being 
thrown? Particularlly one produced by DAQMX would be helpful. 

Copy/paste any error messages.

Original comment by jonathan...@gmail.com on 13 Jul 2011 at 10:18

GoogleCodeExporter commented 9 years ago
Issue is confirmed.

If you use impedance measurement first, then everything works. Otherwise, you 
get this issue. 

Impedance measurer is resolving a routing conflict on the stimulation board for 
some reason. Uhhhh.

Original comment by jonathan...@gmail.com on 2 Apr 2012 at 3:23