Phidgits input reads J thermocouples as K, shows ~200f too high. #61

Closed GoogleCodeExporter closed 8 years ago

GoogleCodeExporter commented 8 years ago
What steps will reproduce the problem?
1. Plug in Phidgits 1048 using J thermocouples
2. Read probe values in  Artisan
3. See problem. 400f reads around 580f. 

What is the expected output? What do you see instead?
Expected output would be lower. I believe Artisan is reading probe values as if 
they were K-type probes. J must have different voltages. 

What version of the product are you using? On what operating system?
Downloaded newest Mac OS release Feb 18 (0.7.4)

Please provide any additional information below.
Cropster offer's a software-side dropdown menu for probe type on each input. I 
can't find a similar option in Artisan. Many many apologies if this feature is 
there and I just couldn't find it. 

Original issue reported on code.google.com by patrickb...@gmail.com on 19 Feb 2014 at 12:32

GoogleCodeExporter commented 8 years ago
Good catch. This feature is missing. Will add as soon as possible and contact 
you for testing.

Original comment by luther.m...@gmail.com on 19 Feb 2014 at 7:07

GoogleCodeExporter commented 8 years ago
[deleted comment]
GoogleCodeExporter commented 8 years ago
Version 0.7.5 now available fixes this. In the Device Assignment dialog (menu 
Config >> Device) there are new settings to set the thermocouple type per 
channel).

Original comment by luther.m...@gmail.com on 11 Apr 2014 at 5:36