Closed kadglass closed 2 weeks ago
yes, the controls for these ccds were tweaked yesterday and we need new darks for them. Klaus
Get Outlook for iOShttps://aka.ms/o0ukef
From: Kelly Douglass @.> Sent: Friday, August 16, 2024 12:30:43 AM To: desihub/nightwatch @.> Cc: Subscribed @.***> Subject: [desihub/nightwatch] Missing data points in CCD QA plots (Issue #426)
Some of the CCDs are missing their data points on the CCD QA plots (e. g. , r1-CD, r4-AB,. . . ). Segev suggested this might be due to missing calibration darks that are raising errors in qproc. — Reply to this email directly, view it on GitHub,
Some of the CCDs are missing their data points on the CCD QA plots (e.g., r1-CD, r4-AB,...). Segev suggested this might be due to missing calibration darks that are raising errors in qproc.
— Reply to this email directly, view it on GitHubhttps://urldefense.com/v3/__https://github.com/desihub/nightwatch/issues/426__;!!KGKeukY!yiHQccBZT3nlG4NlzFwo8gj35_C4YMinpMf6Ythity1Qn6POQgkb_kzKRRi_nuqkYDYS4VYEck2-Iv4RvxxTryKc2A$, or unsubscribehttps://urldefense.com/v3/__https://github.com/notifications/unsubscribe-auth/ABQ5P64VIZIGEMJPIFGLF7LZRV57HAVCNFSM6AAAAABMTLRUOKVHI2DSMVQWIX3LMV43ASLTON2WKOZSGQ3DSNBUGU3DQMQ__;!!KGKeukY!yiHQccBZT3nlG4NlzFwo8gj35_C4YMinpMf6Ythity1Qn6POQgkb_kzKRRi_nuqkYDYS4VYEck2-Iv4RvxzscL37tw$. You are receiving this because you are subscribed to this thread.Message ID: @.***>
Looking further: this is more serious than I thought as several amps (e.g., r1c and r1d) have been missing readnoise, bias, and cosmic rates since the start of April. If this was raised in the nightlog, I missed it before now. Next steps:
Luke pointed out that some of the CCDs are now in 2-amp mode, which could be these "missing" data points. The conversion from 4- to 2-amp mode started in April, so the timing appears to coincide with the above comment. Perhaps all that's needed, then, is an update to Nightwatch to switch the ones in 2-amp mode to only have AB options (or something similar).
Of course. Duh!
Rather than change the plots, I'll probably just add a note to the page so that observers (and I) don't get tripped up on this in the future.