eegkno / ustoolbox

Analysis of the spine from ultrasound images
MIT License
0 stars 0 forks source link

Spatial calibration #10

Closed eegkno closed 6 years ago

eegkno commented 6 years ago

See the section Phantom registration in: http://perk-software.cs.queensu.ca/plus/doc/nightly/user/ApplicationfCal.html

Simulation steps:

  1. Select the Device set named "Saved dataset for both video and tracking (fCal)" from the drop-down box and click Connect
  2. Switch to Freehand calibration toolbox by clicking on its tab on the left side of the application window The video source shows up in the main canvas (By clicking the Edit segmentation parameters button - the "edit-like" button in the line of Segmentation parameters - a dialog appears in which the segmentation parameters can be changed)
  3. Click Start button to start calibration The calibration begins, the segmented points appear as green dots on the image. The green bar on the bottom shows the progress. When it finishes, the result matrix and the errors are displayed in the toolbox.
  4. Show devices checkbox becomes enabled. Clicking on it changes display mode so the tools are displayed in the main canvas instead of the image
  5. The results can be saved by clicking on Save results button as a new XML file
eegkno commented 6 years ago

Things to check when performing spatial calibration: See: http://perk-software.cs.queensu.ca/plus/doc/nightly/user/ApplicationfCal.html

Notes:

Optimize image acquisition to avoid incorrect fiducial segmentations: Fiducial line intersection points should appear as small circular-shaped blobs and the rest of the image should be almost black.

eegkno commented 6 years ago

The wiring should be made with elastic bands to avoid loose wires.

phantom_coord

eegkno commented 6 years ago

In the segmentation parameters:

NOTE: This was the main step to perform the calibration correctly

wiring