The MIO168 module requires 16 calibrations. Adding AFE3 raises that number to 26. We don’t currently have a simple check on which resources are calibrated or not. Also, unlike the power module, it is possible to use it without warning whether the resource is calibrated or not, as well as whether the calibration parameters are enabled or not. This can lead to a situation where the user may unknowingly use the module with a larger measurement error for a long time.
The MIO168 module requires 16 calibrations. Adding AFE3 raises that number to 26. We don’t currently have a simple check on which resources are calibrated or not. Also, unlike the power module, it is possible to use it without warning whether the resource is calibrated or not, as well as whether the calibration parameters are enabled or not. This can lead to a situation where the user may unknowingly use the module with a larger measurement error for a long time.