If trigger (e.g. LIST) wants to be started on CH1, and CH2 has voltage limit set to value lower then the highest from the CH1 list, trigger execution will fail with error "Voltage limit exceeded" despite the fact that on the CH2 trigger is set to FIXed, hence is not affected with trigger.
If trigger (e.g.
LIST
) wants to be started on CH1, and CH2 has voltage limit set to value lower then the highest from the CH1 list, trigger execution will fail with error "Voltage limit exceeded" despite the fact that on the CH2 trigger is set toFIXed
, hence is not affected with trigger.