Unexpected step in measurement #101

Open ronaldDMAS opened 8 months ago

ronaldDMAS commented 8 months ago

Please see the video that shows the unexpected "step" at 8 seconds: https://www.youtube.com/watch?v=1ohqsk5OO6s

My question: is this 'normal' behavior or something that needs a fix? I do not see the behavior on another SA I own.

Ronald

erikkaashoek commented 5 months ago

The step you see if the AGC of the high input reducing gain and thus increasing the noise floor after the strong signal. Suggest to add 20dB more attenuation