ber - met - RH value looks strangly low #21

Closed hagne closed 5 years ago

hagne commented 5 years ago

Shouldn't Bermuda be a rather humid place? Its always below ~35%. Is that an instrument or a data processing/calibration issue?

hagne commented 5 years ago

looks like it was fixed on 11/20/2018

gmd-jw commented 5 years ago

Matt checked the sensor and said that he found a little corrosion that he cleaned off from the connector. However, the corrosion may have been the contact grease we use. Regardless, the data looks reasonable now. A spare sensor has been ordered and will be sent to Bermuda.

gmd-jw commented 5 years ago

Yup. Matt checked the sensor and said that he found a little corrosion that he cleaned off from the connector. However, the corrosion may have been the contact grease we use. Regardless, the data looks reasonable now. Instead of sending them a new humidity element, I have decided to send them a complete new sensor as a spare to set on the shelf in case it acts up again.

Jim

Hagen Telg mailto:notifications@github.com 11/26/2018 17:22

looks like it was fixed on 11/20/2018

— You are receiving this because you were assigned. Reply to this email directly, view it on GitHub https://github.com/hagne/data_baseline/issues/21#issuecomment-441723313, or mute the thread https://github.com/notifications/unsubscribe-auth/Ap1Tqs9_vBd-nlccZXMW7tHiA_IsBOtrks5uzCNCgaJpZM4YMfeR.

gmd-jw commented 5 years ago

I submitted a new issue with a ceilometer. I think it would be good to pull Christian into the GitHub system. He will be managing the ceilometers for SURFRAD. I have already fixed the problem, but I will leave it open so that Christian can close it.