Filter count of 1 not supported by some instruments #67

Closed arnobaer closed 2 years ago

arnobaer commented 2 years ago

A range of Keithley Instruments (2410, 6514, 6517B) do not support a filter (average) count of 1 although the valid range in the User Manuals is defined as 1...100. Setting the value to 1 creates an Invalid Parameter error in the queue.

It was proposed to limit the input range to 2...100.

arnobaer commented 2 years ago

Also set default filter count to 10 according to instrument manuals.