Horos™ is a free, open source medical image viewer. The goal of the Horos Project is to develop a fully functional, 64-bit medical image viewer for OS X. Horos is based upon OsiriX and other open source medical imaging libraries. Horos is made freely available under the GNU Lesser General Public License, Version 3 (LGPL-3.0). Horos is linked against the Grok JPEG 2000 library, for fast viewing of JPEG 2000 images. This library is licensed under the terms of the GNU Affero General Public License.
http://www.horosproject.org
Other
474 stars 142 forks source link

Arthroscopy stills - colors out of range #155

Open raineregle opened 7 years ago

raineregle commented 7 years ago

Our horos' are connected to a Fuji Synapse. When we send stills from our endoscopy cameras (Storz) to Synapse, pictures appear in the right color space. After sending them to Horos, they are in a complete wrong colorspace (mostly green with some pink). I did not find a CLUT to correct it. weasis cannot read it, Osirix 4.1.1-64 same problem, but Escape Medical Viewer can correctly interpret the image. Any idea how to fiddle with horos?

Here a link to one of the misbehaving images: https://www.dropbox.com/s/xsogjk40noligv1/00000001.zip?dl=0

screen shot 2016-12-09 at 12 57 42

buckoO7 commented 6 years ago

I have the same problem with 3D model stills imported from a Fuji Synapse DICOM server. Did you ever find a solution? Thanks

buckoO7 commented 6 years ago

I found a solution to this one:

Osirix Horos Dicom Stills Colours out of range To get rid of the green and pink discolouration, turn off this setting:

Preferences > General > For Color JPEG DICOMs, use JPEG photometric value, instead of DISOM photometric value screenshot 2017-12-04 17 24 05