icub-tech-iit / ventilator-FI5

FI5: Rapidly Manufactured Ventilator System
BSD 3-Clause "New" or "Revised" License
31 stars 4 forks source link

Evaluate pressure sensor for low pressure and O2 compliant #22

Closed MrAndrea closed 4 years ago

MrAndrea commented 4 years ago

This issue to investigate suitable pressure sensors for measure the Air-O2 mixture in the ventilator. The pressure to measure is low (<1bar) and the resolution and sensitivity should be good for the application (0-40mBar). It will be useful investigate also pressure sensor module.

MrAndrea commented 4 years ago

Here something available on Mouser with maximum pressure about 1Bar.

Digital sensors

Analog sensor

Module

simeonedussoni commented 4 years ago

Just for commodity I put a small converting table for pressures:

1mBar = 0.1kPa = 0.0145 PSI

mBar kPa PSI
10 1 0.145
20 2 0.29
40 4 0.58
MrAndrea commented 4 years ago

After 1st April meeting, we look for digital pressure sensor in the 0-100mbar range. Here some proposal.

image

MrAndrea commented 4 years ago

Honeywell offers two useful series:

maggia80 commented 4 years ago

we defined the two sensors for the proto0:

HSCDANN010BGAA5 0-10 BAR

HSCMAND160MD2A5 +-160mbar