Closed jmicrobe closed 6 years ago
Update: This morning I started the thing in debug mode with blank tubes and again got the ~27200 milA readings. Then I removed all tubes and re-started the debug mode (with the channels empty) and I got the expected ~1600 milA readings.
The readings were responsive to testing with my finger obscuring the LED, as well as having light shined in the channel. I did notice that when placing and replacing the tube in the slot there was often a drastic change in milA readings, which sometimes wouldn't recover to the expected ~1600 range.
I'm not sure if I can say that baselining without any tubes in the channels is what specifically corrected the problem. But I do suspect that the movability of the LED/photoresistor pairs in the rack might be it's own issue, and worth reconsidering the housing design for these on the racks.
Kris and I have had this issue a few times, the details are:
The last time this occurred in debug mode I just restarted the ODin about three times and it finally worked - but this hasn't fixed it yet today.